Fridge-tag 2 E

Storage temperature monitor

Contact

We will be pleased to advice you on your individual needs.

Fridge-tag 2 E, newly with 5 years operating time, removes the need for time-consuming handwritten temperature reports. It measures the ambient temperature every minute and immediately issues an alert when your alarm limit is exceeded. The large OK / ALARM indicator on the display enables immediate decision-making. Accurate temperature data from the last 60 days is presented in a PDF which is automatically generated when the device is connected to a PC / Mac, without additional software. The Fridge-tag 2 E is also available with a removable external temperature sensor (measures till –40°C / –40°F).

WHO product qualification

E006/040

Application technical data
Parameters Temperature
Typical application area Storage ambient conditions
Storage cooling conditions (fridges, freezers, displays, etc.)
Markets Pharmaceutical industry
Pharmacies, drugstores, doctors' offices, hospitals, health centers and homes
Food industry, supermarkets and restaurants
Horticulture (plants)
Clinical Trials
Chemical industry
Cosmetic industry
Hardware
External dimensions (L x W x H) 128 x 75 x 19 mm
Size of the LC Display 95 x 27 mm
Weight 96 g
Operating temperature –25°C to +55°C / –13°F to +131°F
Alarm temperature range –20°C to +50°C / –4°F to +122°F, in 0.1°C increments
Alarm and warning limits 2 individually programmable alarm limits (1 upper and 1 lower alarmlimit)
Accuracy of the temperature measurement +/- 0.8°C (–25°C to –10°C) / +/- 1.44°F (–13°F to +14°F)
+/- 0.5°C (–10°C to +40°C) / +/- 0.9°F (+14°F to +104°F)
+/- 0.8°C (+40°C to +55°C) / +/- 1.44°F (+104°F to +131°F)
Memory space 30 day overview on the display / up to 60 days' overview in the PDF report
Protection class IP64
Operating time Up to 5 years - expiry date printed on label
Battery life Up to 5 1/2 years (1/2 year storage / 5 years useful life). The battery indicator on the display provides information on the remaining lifetime.
Activation Pressing the SET and READ button for at least 3 seconds
Stop No stop function
Configuration Pre-set by Berlinger & Co. AG or self-configuration
Software
Format of the Original document PDF and ASCII file (it can also be imported by many ERP or data management systems)
Software for verifying the PDF and ASCII file Berlinger Verifier (www.berlinger.com/verifier) / JAVA™ / The software verifies the digital signature of the ASCII and PDF files. FDA regulation 21 CFR Part 11 compliant. Software is compliant with GAMP 5 requirements.
Compatibility with Berlinger SmartView No
NIST / ILAC calibration certificate On request

Accessories