Product Overview

Temperature Monitoring Devices

Fridge-tag 2 E

Storage temperature monitor

Fridge-tag 2 E
Fridge-tag 2 E
Fridge-tag 2 E
Fridge-tag 2 E
Contact

We will be pleased to advice you on your individual needs.

Fridge-tag 2 E, newly with 5 years operating time, removes the need for time-consuming handwritten temperature reports. It measures the ambient temperature every minute and immediately issues an alert when your alarm limit is exceeded. The large OK / ALARM indicator on the display enables immediate decision-making. Accurate temperature data from the last 60 days is presented in a PDF which is automatically generated when the device is connected to a PC / Mac, without additional software. The Fridge-tag 2 E is also available with a removable external temperature sensor (measures till –40°C / –40°F).

See also: Brochure Fridge-tag devices

WHO product qualification

E006/040

Application technical data
ParametersTemperature
Typical application area Storage ambient conditions
Storage cooling conditions (fridges, freezers, displays, etc.)
Markets Pharmaceutical industry
Pharmacies, drugstores, doctors' offices, hospitals, health centers and homes
Clinical Trials
Chemical industry
Hardware
External dimensions (L x W x H)128 x 75 x 19 mm
Size of the LC Display95 x 27 mm
Weight96 g
Operating temperature–25°C to +55°C / –13°F to +131°F
Alarm temperature range–20°C to +50°C / –4°F to +122°F, in 0.1°C increments
Alarm and warning limits2 individually programmable alarm limits (1 upper and 1 lower alarmlimit)
Accuracy of the temperature measurement +/- 0.8°C (–25°C to –10°C) / +/- 1.44°F (–13°F to +14°F)
+/- 0.5°C (–10°C to +40°C) / +/- 0.9°F (+14°F to +104°F)
+/- 0.8°C (+40°C to +55°C) / +/- 1.44°F (+104°F to +131°F)
Memory space30 day overview on the display / up to 60 days' overview in the PDF report
Protection classIP64
Operating timeUp to 5 years - expiry date printed on label
Battery lifeUp to 5 1/2 years (1/2 year storage / 5 years useful life). The battery indicator on the display provides information on the remaining lifetime.
Activation Pressing the SET and READ button for at least 3 seconds
Stop No stop function
Configuration Pre-set by Berlinger & Co. AG or self-configuration
Software
Format of the Original document PDF and ASCII file (it can also be imported by many ERP or data management systems)
Software for verifying the PDF and ASCII fileBerlinger Verifier (www.berlinger.com/verifier) / JAVA™ / The software verifies the digital signature of the ASCII and PDF files.
Compatibility with Berlinger SmartView No
NIST / ILAC calibration certificate On request

Accessories