Product Overview

Temperature Monitoring Devices

Q-tag CLm doc L

Multi-alarm shipping temperature logger

Q-tag CLm doc L
Q-tag CLm doc L
Q-tag CLm doc L
Q-tag CLm doc L
Q-tag CLm doc L
Contact

We will be pleased to advice you on your individual needs.

The Q-tag CLm doc L is a single-use data logger with a USB interface for data analysis on a computer. When connected to a PC / Mac, the logger automatically generates a secure PDF/A report showing the measured data including the temperature curve.

  • Runtime up to 365 days thanks to a large memory capacity of 38,000 measurement points.
  • Easy to read out with √ or Χ
  • User-friendly, reliable and high-precision

Learn more: Brochure Q-tag CLm doc Devices

WHO product qualification

E006/030

Application technical data
ParametersTemperature
Typical application area Shipment Temperature Logger
Markets Pharmaceutical industry
Pharmacies, drugstores, doctors' offices, hospitals, health centers and homes
Clinical Trials
Hardware
External dimensions (L x W x H)94 x 54 x 20 mm
Size of the LC Display25 x 19 mm
Weight35 g
Operating temperature–30°C to +60°C / –22°F to +140°F
Alarm temperature range–25°C to +55°C / –13°F to +131°F, in 0.1°C increments
Alarm and warning limitsUp to 5 individually programmable alarm limits
Logging intervalAdaptable 1 to 60 minutes
Accuracy of the temperature measurement +/- 0.8°C (–30°C to –10°C) / +/- 1.44°F (–22°F to +14°F)
+/- 0.5°C (–10°C to +40°C) / +/- 0.9°F (+14°F to +104°F)
+/- 0.8°C (+40°C to +60°C) / +/- 1.44°F (+104°F to +140°F)
Memory space38’000 points
Protection classIP65
Operating timeUp to 2 years (1 year storage / 1 year useful life). Battery warning: display blinking
Activation Start button
Stop Stop button
Configuration Pre-set by Berlinger & Co. AG
Software
Format of the Original document PDF and ASCII file (it can also be imported by many ERP or data management systems)
Software for verifying the PDF and ASCII fileBerlinger Verifier (www.berlinger.com/verifier) / JAVA™ / The software verifies the digital signature of the ASCII and PDF files.
Compatibility with Berlinger SmartView Yes
NIST / ILAC calibration certificate On request